STD Model
Internal Size |
External Size |
Capacity |
Shelves provided |
Standard |
W x D x H cm |
W x D x H cm |
Cu.ft, Liters |
(S.S. trays) |
Model No. |
45 x 45 x 60 |
65 x 114 x 130 |
4 Cu ft., 120 Liter |
2 |
SSE - 4 |
60 x 60 x 60 |
76 x 121 x 130 |
8 Cu ft., 200 Liter |
2 |
SSE – 8 |
60 x 60 x 90 |
76 x 121 x 165 |
12 Cu ft., 324 Liter |
3 |
SSE – 12 |
60 x 60 x 125 |
76 x 121 x 199 |
16 Cu ft., 450 Liter |
4 |
SSE – 16 |
60 x 80 x 125 |
76 x 141 x 199 |
21 Cu ft., 600 Liter |
5 |
SSE - 21 |
Specification:
For conditions as per ICH |
25°C – 60% RH 40°C – 75%RH 30°C – 65%RH |
Construction |
Double wall with Poly Urethene (PUF) insulation provided with door having locking arrangement and inner glass viewing door. |
Temperature range |
10.0 °C to 60.0 °C. |
Temperature resolution |
0.1 °C |
Temperature accuracy |
+0.2° C. |
Temperature uniformity |
+1°C |
Temperature Control |
Microprocessor based PID Control. With Auto Tune CE Marked |
Temperature sensor |
PT100 RTD type Class ‘A’, Made in Germany |
Temperature sensor accuracy |
+0.25 °C |
Humidity range |
40% to 90% RH |
Humidity resolution |
1% |
Humidity accuracy |
+2% RH |
Humidity uniformity |
+3 % RH |
Humidity sensor |
Make HUMIREL made in France, Capacitance type High reliability and long term stability Calibrated within +2% RH @ 55% RH Humidity measuring range 0 to 100% Not affected by water immersion |
Cooling: |
CFC Free Copeland Make compressor utilizing R 134A eco friendly refrigerant, with condenser, motor, relay complete unit Copeland Make |
Steam Injection |
Boiler with reservoir made of thick stainless steel, with heater, water inlet control, low water level safeguard and insulation. |
Heating |
‘U’ Shaped Nichrome Wire heater in SS Sheathing |
Air circulation |
Flange motor with impeller / blower. |
Insulation |
Poly Urethane (PUF) |
Chamber illumination |
Fluorescent light with door switch. |
Trays |
SS wire meshes heavy duty |
Feet |
Castor wheels. |
Electrical |
230V / 15A / 50 Hz. |
Safety Features:
Temperature |
High temperature safety cut off. |
Boiler heater |
Low water level boiler cut off. |
Compressor |
Electrical overload cutoff relay for compressor. |
Electrical |
Time delay for compressor switch ON. |
PRE INSTALLATION MANDATORY REQUIREMENT : (Guarantee void incase of non compliance) |
a) Distilled water OR DM water continuous supply with tap for boiler input of ½” BSP at 3 feet height from ground level. |
b) Room temperature around machine preferably at 25°C with air conditioning or a well ventilated room with exhaust fan. However surrounding temperature should not exceed 30°C. |
c) Stabilized Input Voltage of 230V AC 20 Amps. Use of Servo Controlled Stabilizer is recommended. |
d) Water drains line with ½” nozzle to be provided at floor level. |
Calibration & Validation
|
|
Calibration |
Temperature sensor probe calibration with traceability to ERTL Humidity sensor calibration with traceability to ERTL Temperature controller calibration with traceability to ERTL Humidity controller calibration with traceability to ERTL |
Validation |
IQ, OQ and PQ documentation and protocols shall be provided with equipment. Documentation prepared / developed for Scientific International by Premier Validation Ltd – Ireland, one of Europe’s leading validation consultants. |