Helpline No.: +91-9971882305

     
OUR PRODUCT CATEGORIES
 
Contact Us
Mr. Saurav
C-238, Yamuna Vihar, Delhi-110053, India
Mobile : +91-9971882305
Email :info@sunscientific.co.in
Humidity / Stability Test Chamber
 
  • Precise Hot Air & Humidity
  • Better Mineral / Glasswool / Puf Insulation Keep the surface Temperature comfortable & Avoid Heat & Humidity Loss.
  • Silent Hot Air Blower, Unique design of Air Circulation & Humidity circulation & improved Temperature / Humidity distribution.
  • Unique design of Ventilation.
  • Polish 304 grade of S. Steel interior Corrosion resistant long life operation
  • Kanthal A-1 Tubular heater is fitted for better circulation
  • Safety controller are provided with Digital Temp. Controller Cum Indicator
  • Kanthal A-1 steel Tubular heater J shape is fitted in the steam Bath for long life & better result.
  • Full feature with Temp. Controller i.e. Thermostate Bi-Metalic Capillary type for both Dry & Wet Temp. Control or Digital Temp. Controller Cum Indicator & Microprocessor Based PID Temp. Controller Cum Indicator
  • Choose among the models with window front glass door or inner Perspex / Acrylic sheet Door with full view to observe the Test process inside the chamber while keep Temperature uninterrupted.
  • Angle Iron frame with castor wheel for easy movement (for Refrigerated model )
  • Fitted with 02 Nos. Fan for cooling & heating (For Refrigerated Model)
  • Filled with 02 Nos. Circulation motor for water injection in routine model inside the control unit & Chamber.
  • Humidistate is placed inside the chamber to control requested humidity.
  • Humidity can be controlled by Digital Temp. & Humidity (Both automatically on customer requirement) •Wet Temp. is controlled by Bi-Metalic Thermostate Connected to Water Bath placed in the control unit having water level. (Nozzle for Water inlet & outlet & overflow)
Applications :

Tissue Culture, Pharmaceuticals, Agriculture, Cable Industry, Paper , PVC & Cement Industries Electrical Industries.

Technical Parameters :

Temperature Range :

Ambient to 45 °C

Temperature Stability :

+-1°C

Housing Material :

CRC / Mild Steel Sheet duly Powder Coated

Power Consumption :

220/230V, AC +- 10% 50Hz Single Phase

Safety Function :

Water Level & Low water level cut off alarm (On customer request)

Optional :

RS 232 Communication Interface

Duty Cycle :

Continuous

Interior :

S.S. 304 Polished Sheet

Humidity Range:

Ambient to 95% +-3% (For Routine Model)

Temp. Range :

10 - 50°C (For Cooling Refrigerated Model)

Humidity Range :

55 - 92% +-3% (For Cooling Refrigerated Model)

Compressor :

Kirloskar/ Tecumseh As per model requirement (For Cooling below Ambient)

Voltage Stabilizer :

Required for Cooling (Refrigerated Model as per size/ model)

Refrigerant :

R 134/22 (For Cooling Model)

Optional :

  • Digital Timer
  • Wet & Dry Thermometer for Calibration

Precaution :

Don't use the Instrument without water