For measurement of pH, conductivity, mv, TDS, Salinity, D.O. & temperature 16x2 Alphanumeric LCD display, Battery cum mains operated with all accessories fitted in an briefcase. Data storage facility up to 700 samples and external printer attachment facility available.
Technical Specifications:
General
Display: 16 characters x 2 lines alpha Numeric BL LCD
Display Format: Temperature and parameter Simultaneously
Controls: By soft keys Menu Driven
Storage: Up to 700 samples
Power supply: 230V + 10% 50 Hz OR 12V battery
pH
Range: 0 – 14 pH
Resolution: 0.01pH
Accuracy: + 0.01 pH (relative)
Temp. Comp.: Automatic with Pt 100 Sensor
Buffer Recognition: 4.00, 7.00, 9.20 (Automatic)
Calibration: Automatic & Manual
Dissolved Oxygen
Range: 0 to 20 ppm
Resolution: 0.1 ppm
Calibration: Automatic
Accuracy: + 0.2 ppm + 1 digit
Temp. Comp.: Automatic with Pt 100 Sensor
Pressure Comp.: 0 – 999 mm of Hg (Manual)
Salinity Correction: 0 – 20 mg/l
mV / O.R.P.
Range: -2000 to +2000mV
Resolution: 1mV
Calibration: Manual
Accuracy: + 0.1% + 1 digit
Temperature:
Range: -100o to +200oC
Resolution: 0.1oC
Accuracy: + 0.2% + 1 digit
Input Pt.: 100
Calibration: Manual
Conductivity:
Range: 0 – 200.0 mS/cm
Range: Selection Automatic
Resolution: 0.01 uS/cm
Accuracy: + 0.5%
Temp. Comp.: Automatic with Pt 100 Sensor
Ref. Temp.: 25 degree C
Temp.: Co - efficient 0 – 4% adjustable
Cell constant: 0.05 to 2.0 adjustable
T.D.S.
Range: 0 – 200 ppt
Range Selection: Automatic
Resolution: 0.1 ppm
Accuracy: + 0.5%
TDS factor: 0.5 by default (adjustable)
Salinity
Range: 0 – 50 ppt
Resolution: 0.1 ppt
Temp. Comp.: Automatic
Accuracy: + 0.5%
Salinity factor: 0.75 by default (adjustable)
Option:
RS 232 output available for interfacing to computer / printer
Appearance and specifications subject to change due to continuous development
Accessories:
- pH Electrode
- Two Buffer solutions
- Conductivity / TDS / Salinity Cell
- DO Probe, D.O. Membranes, KCL
- Temp Probe
- Operation manual
- ORP Electrode AT EXTRA COST